Spectro Wipe |
Options
- Dual Detector Interface
- Laser printing capability
|
|
The SpectroWipeTM Wipe Test Counter is a compact radiation measurement device for determining levels of contamination in Nuclear Medicine laboratories and other environments where radioisotopes are used.
It includes a 1024-channel Multi-Channel Analyzer for precision peak detection and isotope identification by gamma spectroscopy.
Counting data, spectra, and activity data are displayed on a large, bright LCD display with adjustable contrast. With a maximum count rate of 100,000cps and a high efficiency NaI detector, the SpectroWipe performs wipe tests in a minimal amount of time.The easy-to-read display and menu-driven user interface make wipe
test counting easy and efficient, streamlining quality control and safety procedures, while reducing paperwork and manual record keeping.
The spectrum is displayed while it is being acquired, along with
the count rate, total counts, and time remaining. Background is
automatically subtracted. When the counting is complete, the
detected peaks are displayed, the identified nuclides are listed, and
the calculated activities are displayed for each. The total activity is
also displayed, automatically converted to the desired units (DPM,
μCi, or Bq), and exceeded trigger levels are indicated. |
|
|
Wipe Test Counting System
with 1024-channel MCA
Features
- Meets NRC regulations
- NaI drilled-well crystal detector with built-in preamplifier
- 1024-channel MCA
- Precision high voltage
- Preset count or preset time mode
- Automatic background subtraction
- Precision peak detection
- Automatic Isotope identification
- Nuclide library stores up to 50 nuclides
- Up to 20 locations
- Data storage and archiving
- Prints reports, spectra, data
- RS-232 port for data input/output
- Large, backlit LCD display
- Easy to use, menu-driven interface
- C-US certification by TUV Rheinland
|
|
Wipe test, background, and calibration data are archived, and
reports can be printed on a serial or parallel printer, or the data can
be exported via RS-232 to a computer, in spreadsheet format, and
then printed from the computer.
Settings are stored by location, including trigger level, energy range,
nuclides, LLD, and ULD. Up to 20 numbered and named locations
can be stored. The system contains a database of radionuclides
which stores the half-life, peaks, ROI, and efficiency for each.
Data is saved in 128kB flash memory. Available reports include
monthly, weekly, and daily activity by location.
Tested to EN61010-1 and equivalent UL/CSA standards for CAN/USA certification by TUV Rheinland, and
tested to EMC standard EN61326 for CE compliance.
AccuSync Medical Research Corporation is
an ISO9001 and ISO13485 registered
designer and manufacturer of medical
devices and related instruments in nuclear
medicine. |
|
Specifications |
Detector |
|
Drilled-Well Crystal:
|
NaI scintillator shielded by 0.847” lead.
Crystal dimensions: φ1.75”x2”. Drilled well: φ.656”x1.75”. |
Preamplifier:
|
The detector has a built-in, charge sensitive preamplifier with positive
output. The required power is ± 5V @ 2 mA. |
Multi-Channel Analyzer: |
|
Linear Amplifier:
|
1600keV or 2000keV energy range. |
Max Count Rate:
|
100,000 CPS |
Pulse Pair Resolution:
|
1μs |
LLD:
|
The low level discriminator can be set to any level up to 100% of the selected range in 10 keV steps. |
ULD:
|
The upper level discriminator can be set to any level up to 100% of the selected range in 10 keV steps. |
Number of Channels:
|
1024 |
Calibration:
|
Automatic; standard source required. |
Peak Detection:
|
Automatic; with automatic nuclide identification of up to 50 nuclides. |
Conversion Time:
|
10μs |
High Voltage: |
500-1000V, electronically stabilized. Output Current: 500uA
Temp stability: 50ppm Ripple: <25mV peak-to-peak |
Display: |
Menu-driven graphic display.
320x240 Monochrome LCD with backlight. |
Keypad: |
6 membrane switches for menu-driven graphic user interface. |
Input/Output: |
Preamp input in 6 pin miniature connector .
Bias Voltage SHV connector.
RS-232 ports for printing and exporting data to computer.
TTL pulse output. |
Clock: |
Battery backed-up clock/calendar keeps current time and date. |
Memory: |
128k EEPROM for data storage. Storage capacity for spectra, location data,
wipe test results, nuclide data, calibration and background logs. |
Printer: |
Parallel printer capability with converter for HP LaserJet and PCL-compatible
printers. |
Power: |
85-265V~, 47-440Hz, less than 5W |
Dimensions: |
|
Detector:
|
Diameter 4.00”, Height 9.25”, Weight 30 lbs. |
MCA Unit:
|
Width 7.00”, Length 7.09”, Height 4.25”, Weight 3.5 lbs. |
|