Products » ECG Triggers » Spectro Analyzer
 
Spectro Analyzer

Options

  • Single- or Dual-Channel Analyzer
  • 1024-channel Multi-Channel Analyzer
  • Laserprinting capability
  • Well detector with 2” lead for PET
 

The SpectroAnalyzerTM is a compact single or dual-channel counting instrument used for gamma counting, and is also available with a 1024-channel MCA. It contains a precision pulse height analyzer and high voltage supply, both microprocessor controlled.

The pulse height analyzer setting is accomplished through amenu and selects an energy region of interest. When the proper high voltage is applied to the detector, the PHA is calibrated directly in keV. Optional spectrum/MCAincludes auto-calibration.

Output signals from the pulse height analyzer are totaled by the scaler, which can operate in either a preset count or preset time mode, and background count is automatically subtracted. Efficiency can bemeasured or entered, and activitycan be displayed in disintegrations perminute (DPM).

 

Single / Dual-Channel Analyzer
Multi-Channel Analyzer

Features

  • 7-decade scaler
  • Digital rate meter with a range of 0-10million counts per minute
  • Precision pulse height analyzer
  • Variable amplifier for energy range
  • Preset count or preset time mode
  • Automatic background subtraction
  • Data can be displayed in DPM
  • Precision high voltage
  • NaI drilled-well crystal detector with built-in preamplifier
  • Up to 20 locations
  • Data storage and archiving
  • Prints reports, spectra, data
  • RS-232 port for data input/output
  • Large, backlit LCD display
  • Easy to use,menu-driven interface
  • CAN/USA certification by TUV Rheinland
  • CE certified
 

Aserial computer interface (RS-232) for input and output allows a computer to change settings, and count data can be transferred to the computer for printing or saving to disk.

CE

Tested to EN61010-1 and equivalent UL/CSAstandards for CAN/USA certification by TUV Rheinland, and tested to EMC standard EN61326 for CE conformity.

CE

AccuSyncMedicalResearchCorporation is an ISO9001 and ISO13485 registered designer and manufacturer of medical devices and related instruments in nuclear medicine.

Specifications

Resolving Time:

Onemicrosecond pulse pair. The overall resolving time of the scaler is limited by the input circuits and the necessary integration time in the detector for adequate pulse height resolution.

Time:

Themicroprocessor-controlled timer indicates elapsed time up to 1000minutes.

PresetTime:

Time can be preset to any number up to 1000 minutes using themenu.

PresetCount:

Number of counts from100 to 10million can be preset by using themenu.

Background:

Automatic subtraction of background count.

Digital Rate Meter:

Indicates 0-10 million counts per minute with automatic change of range.

Pulse Height:

Five energy ranges from 500 to 2000 keV in steps of 500 keV are provided.

Window Setting:

1) Upper and lower leveldiscriminator independently controlled from 0 to 100%.
2) % window. Lower level is set as the reference in keV’s. Windowis set as the% of this reference.
3) Symmetricalwindow. Lower level is set as the reference in keV’s. Windowis set symmetrical about this reference.

Multi-ChannelAnalyzer:

Linear Amplifier:
Max count rate:
LLD:
ULD:
No. of Channels:
Calibration:
Conversion time:

 

The variable gain amplifier covers the energy range from500keV to 2000keV in 500keV steps.
100,000 cps
Lowlevel discriminator can be set to any level up to 100% of the selected range in 10keVsteps.
Upper level discriminator can be set to any levelup to 100% of the selected range in 10keVsteps.
1024
Automatic; standard source required.
10us

Display:

Backlighted LCD display provides indication of accumulated counts, elapsed time, rate of incoming pulses, lower and upper level discriminator values, and high voltage setting.

Keypad:

6membrane switches.

PHA Options:

The PHA consists of one or two independent channels with capability of one or two detectors.
The following options are available:
Option 1: Single Channel PHA for single detector with or without 1024-channel MCA
Option 2: Dual Channel PHA for single detector without 1024-channel MCA

High Voltage:

500-1000V, electronically stabilized. Output Current: 500uA
Temp stability: 50ppm Ripple: <25mVpeak-to-peak

RS232 Output:

Count data, time, lower and upper level energy level settings, background counts perminute.

Input Sensitivity:

One millivolt.

Power:

90-240V~, 47-440 Hz, less than 5W

Battery:

Optional 9.6V NiMHrechargeable batterywith 7 hrs. approximate running time.

Detector:

Ø1.75”x2” NaI crystal shielded by 0.847” lead. Optional 2” for PET. Drilled well: Ø.656”x1.75”
Preamplifier: The detector has a built-in, charge sensitive preamplifier with positive output.
The required power is +/-5V@2mA.

Dimensions:

Detector (DW1000):  Diameter (base) 5.50”, Height 9.25”, Weight 30 lbs.
Detector (DW1000P for PET): Base: 8” square, Height 12.25”, Weight 150 lbs.
MCA unit: Width 7.00”, Length 7.09”, Height 4.25”, Weight 3.5 lbs. (w/ battery 4.5lbs.)